Shunt Valves — How They Work

Pterion Prep · Shunts

Everything about shunt valves, in one place

Start here for the physics and device types, then look up any valve's settings in mmH₂O and mmHg, convert a setting between brands, or identify a valve on X-ray. Every number on these pages is taken from the manufacturer's instructions for use, brochure or an FDA record — sources are linked on each valve card.

Educational reference only. Device behaviour depends on the exact model, configuration, catheters and patient. Confirm against the implant record, the current IFU and your attending before changing any setting.

The 60-second version

1. Pressure units

Unit In mmH₂O Used by
1 mmH₂O 1 Codman/Integra, Medtronic, Sophysa valve settings
1 cmH₂O 10 Miethke / B. Braun valves, manometers, LP opening pressure
1 mmHg 13.595 ICP monitors, physiology texts
1 kPa 101.97 SI literature

Manufacturer shorthand: Miethke uses 1 mmHg ≈ 1.36 cmH₂O (and 1 cmH₂O ≈ 0.74 mmHg); Sophysa states 1 mmH₂O = 9.807 Pa = 0.074 mmHg.

mmH₂O cmH₂O mmHg
50 5 3.7
100 10 7.4
136 13.6 10
150 15 11.0
200 20 14.7
204 20.4 15
272 27.2 20
400 40 29.4

Use the quick unit converter for any other value.

2. What “opening pressure” really means

It is a pressure-flow curve, not a single number. Each valve passes more CSF as the pressure difference rises, so manufacturers quote the pressure at a reference flow — and they do not all use the same one:

Manufacturer / valve How the labelled pressure is defined Stated tolerance
Codman HAKIM programmable Nominal setting (30–200 mmH₂O, 10-mm steps) ±14 (30–120), ±19 (130–200) mmH₂O
Codman CERTAS Plus Average across 5–50 mL/h ±20 (1–3), ±25 (4), ±35 (5–7) mmH₂O
Codman HAKIM Precision (fixed) Labelled operating pressure ±10 mmH₂O
Medtronic Strata, Delta, CSF-Flow Control Median range from 5 to 50 mL/h Strata NSC: ±15–25 mmH₂O
Sophysa Polaris, Sophy At 10 mL/h, valve alone −10 / +15 mmH₂O
Miethke gravitational units At 5 mL/h (≈1–2 cmH₂O higher at 20 mL/h) —

Catheters add resistance. Integra’s L-P IFU shows the gap: a HAKIM valve set at 30 mmH₂O measured 61 mmH₂O with a 120 cm peritoneal catheter at 20 mL/h, and setting 200 measured 229. Sophysa labels the valve alone and notes the catheters add their own resistance; Medtronic’s small-lumen peritoneal catheter limits flow about six-fold versus a standard one at 20 mL/h.

So “the same number” on two brands is only approximately the same pressure — which is why the converter reports the closest setting, the difference, and the next setting above and below.

3. Posture, hydrostatics and the siphon effect

Miethke’s IFUs give the cleanest model. With PL the valve opening pressure lying, PS the opening pressure upright, PH the hydrostatic column from the ventricles (foramen of Monro) down to the distal cavity, and PB the abdominal pressure:

In an upright adult the column is tens of centimetres (42 cm in Miethke’s worked example), far above any lying setting — so a differential-pressure valve alone lets IVP fall well below zero when the patient stands: overdrainage, with slit ventricles, low-pressure headache, hygroma or subdural haematoma. Raising the setting to stop it simply under-drains the patient when lying. The fix is a device that adds resistance only upright, which is what the next two sections are about.

Medtronic simulates the upright patient by testing at −50 cmH₂O hydrostatic pressure; Miethke aims for a slightly negative standing IVP (the DualSwitch valve targets −5 to +5 cmH₂O).

4. Valve mechanisms

Mechanism How it works Examples
Ball-in-cone (spring-loaded) A ball (synthetic ruby / sapphire) is pressed into a conical seat by a spring; the spring force sets the opening pressure Codman HAKIM, CERTAS Plus, HAKIM Precision; Sophysa Polaris, Sophy; Miethke DP units; Integra DP (two in series); Medtronic Strata/StrataMR II ball-and-cone
Membrane / diaphragm A flexible membrane seals an outlet until pressure lifts it Medtronic Delta, CSF-Flow Control (non-metallic)
Slit valve Silicone slits open under pressure Historic Holter / Spitz-Holter; distal slit “closing-pressure” catheters (Medtronic: low 0.5–5.4, medium 5.5–9.4, high 9.5–14.4 cmH₂O — not flow-control valves)
Flow-regulating Variable resistance keeps flow near CSF production instead of holding a set pressure Integra OSV II (18–30 mL/h; Low Flow model 8–17 mL/h)
Adjustable flow resistance A rotor sets the length of a flow channel; can switch the shunt off Miethke M.flow (used with M.blue)

How programmable valves change setting: HAKIM — a codified magnetic signal drives a stepper motor that turns a cam under the spring. CERTAS Plus — a magnetically rotated construct. Strata — a magnet-driven mechanism read with hand tools or StrataVarius. Polaris — a rotor held by a magnetic lock (two opposed micro-magnets a one-directional field cannot release). Sophy — a magnetic rotor with no lock. Miethke — a rotor held by a brake that releases only when the adjustment tool presses on the valve (M.blue adds an “Active-Lock”).

5. Siphon-control (“anti-siphon”) devices

These respond to the suction or flow surge of upright posture rather than to posture itself.

6. Gravitational (“anti-gravity”) units — built-in and add-on

Principle: a heavy tantalum ball rests on a ball-and-cone seat. Lying flat, the ball rolls off the seat and adds no resistance; upright, its weight must be lifted, adding a fixed opening pressure. Intermediate angles add a graded amount. They must be implanted parallel to the body axis.

Device Built-in or add-on Gravitational pressure Adjustable?
proGAV 2.0 (Miethke) Built-in (SHUNTASSISTANT 2.0) on an adjustable 0–20 cmH₂O DP unit 10 / 15 / 20 / 25 / 30 / 35 cmH₂O (or none) DP yes; gravity fixed
proGAV (1st gen) Built-in on an adjustable 0–20 cmH₂O DP unit 10–35 cmH₂O (or none) DP yes; gravity fixed
M.blue (Miethke) Built-in — one valve: fixed DP unit 0 / 5 / 10 / 15 cmH₂O 0–40 cmH₂O, continuously Gravity yes; DP fixed
M.blue plus Built-in: proGAV 2.0 DP unit + M.blue gravitational unit 0–40 cmH₂O Both
GAV / GAV 2.0 / paediGAV Built-in fixed pairs e.g., GAV 2.0 5/20 … 10/30 cmH₂O No
DualSwitch valve Built-in: separate lying (5–16) and upright (30–50 cmH₂O) chambers Switch, not additive No
SHUNTASSISTANT 2.0 Add-on, in series below any DP valve 10–35 cmH₂O No
SHUNTASSISTANT / paediSA Add-on 15–35 / 10 cmH₂O No
proSA Add-on 0–40 cmH₂O (preset 20) Yes
SiphonX (Sophysa) Add-on downstream of a Polaris / Sophy (pre-attached on “-SX” models) +200 mmH₂O fully upright No
Gravity Compensating Accessory (Integra) Add-on, anywhere on distal tubing 2–4 steel balls; value not published No

Is the M.blue an attachment? No — the M.blue is a complete valve whose gravitational unit is the adjustable part. The true attachments are the SHUNTASSISTANT family, proSA, SiphonX and the Integra GCA, which you add in series to a separate differential-pressure valve.

Total upright opening pressure of a gravitational system = DP unit + gravitational unit. Example (Miethke): DP set to 10 cmH₂O with a 25 cmH₂O gravitational unit → 10 cmH₂O lying, 35 cmH₂O upright.

7. MRI and programmable valves

Valve MRI effect on the setting After MRI
CERTAS Plus Resists unintended change at 3 T Confirm the setting after MRI (Integra).
Hakim Programmable (CHPV) Up to 3 T will not damage it but MAY CHANGE THE SETTING Always confirm the setting after MRI (X-ray or programmer). Repeated 3 T exposure may demagnetize the rotor and reduce programmability.
Strata II (with Delta chamber) 3 T will not damage the valve but may change the setting Check the performance level immediately after MRI (Medtronic).
Strata NSC (no siphon control) 3 T will not damage the valve but may change the setting Check the performance level immediately after MRI.
StrataMR II Designed to withstand unintended change up to 3.0 T Medtronic still advises verifying the performance level after MRI. MR Conditional 1.5/3 T (max spatial gradient 1900 G/cm).
Polaris (SPV) MRI-stable (magnetic lock); MR Conditional 1.5/3 T Sophysa still recommends checking the setting before and after MRI; valve rotation during a 3 T exposure can change it.
Sophy Mini SM8 MR Conditional 1.5/3 T but NO magnetic lock — high risk of accidental change Check the setting before AND after every MRI (adjustment kit or X-ray).
Sophy SM3 No magnetic lock Check the setting before and after MRI.
proGAV 2.0 MR Conditional ≤3 T — no setting change Miethke: verification after MRI is not required.
proGAV (1st generation) MR Conditional ≤3 T — no setting change Miethke: verification after MRI is not required.
M.blue MR Conditional ≤3 T — no setting change; Active-Lock Miethke: no X-ray verification after MRI needed.
M.blue plus MR Conditional ≤3 T — no setting change Miethke: no verification after MRI needed.

Fixed valves have nothing to reprogram, but every valve makes some artifact; the non-metallic Medtronic Delta and CSF-Flow Control valves make essentially none. Outside the scanner, strong everyday magnets can also move a setting — Integra warns that magnets over 80 gauss (household or loudspeaker magnets) near a HAKIM valve can change it, while Miethke’s braked rotors and Sophysa’s Polaris lock are designed to resist them.

8. Adjusting a setting safely — manufacturer rules

9. Manufacturer starting-pressure recommendations (non-binding)

These are the manufacturers’ own suggestions — not clinical guidelines — and Miethke stresses they must be individualised (lower the gravitational unit for low mobility or high BMI; no gravitational unit for bedridden patients).

M.blue (DP unit 5 cmH₂O for all; upright total in brackets):

Patient Gravitational unit (cmH₂O) Upright total
Newborns and children < 5 y 20 25
Children ≥ 5 y 25 30
Adults (< 1.60 m / > 1.80 m) 25 (20 / 30) 30 (25 / 35)
Adults > 65 y (< 1.60 m / > 1.80 m) 20 (15 / 25) 25 (20 / 30)

proGAV: DP 5 cmH₂O standard for NPH; gravitational unit 20 cmH₂O (children ≤5 y, adults >60 y) or 25 (children >5 y, adults ≤60 y); −5 if shorter than 1.60 m, +5 if taller than 1.80 m.

paediGAV: <6 months 4/24; 6 months–5 years 9/24; >5 years 9/29 cmH₂O (lying/upright).

Sizing an add-on gravitational unit (Miethke method): choose the unit so that DP valve + gravitational unit ≈ (diaphragm-to-foramen-of-Monro distance) − 5 cm. Worked example from the IFU: 42 cm distance, DP valve 13 cmH₂O → 42 − 13 − 5 = 24 → select the 25 cmH₂O unit → standing IVP ≈ 25 + 13 − 42 = −4 cmH₂O. The DualSwitch valve uses the same distance-minus-5-cm rule for its upright chamber.

10. Recalls and safety notices

11. Using the converter well

  1. Choose the source valve and setting (or type a pressure) and the posture you care about.
  2. Read the result in all four units, plus the lying and upright opening pressures for that setting.
  3. Look at the closest equivalent on the target valve and its neighbours above and below — then pick using clinical context, not just the smallest difference.
  4. Remember the caveats: different reference flows, ±10–35 mmH₂O tolerances, catheter resistance, and fundamentally different siphon and gravity behaviour.

Open the converter →